Journal article
Authors list: Richter, Manuel J.; Harutyunova, Satenik; Bollmann, Tom; Classen, Simon; Fuge, Jan; Gall, Henning; Gerhardt, Felix; Ghofrani, Hossein A.; Gunkel, Hartmut; Gruenig, Ekkehard; Halank, Michael; Heine, Alexander; Klose, Hans; Lange, Tobias J.; Neurohr, Claus; Nickolaus, Kai; Opitz, Christian F.; Rosenkranz, Stephan; Seyfarth, Hans-Juergen; Tello, Khodr; Ewert, Ralf; Olsson, Karen M.
Publication year: 2020
Journal: Pulmonary Circulation
Volume number: 10
Issue number: 1
ISSN: 2045-8932
eISSN: 2045-8940
Open access status: Gold
DOI Link: https://doi.org/10.1177/2045894020910136
Publisher: Wiley
Abstract:
Implantable infusion pumps might improve the convenience and safety of intravenous treprostinil for pulmonary arterial hypertension. The LENUS Pro (R) pump (approved in Europe) has a fixed flow rate. Based on 126 pumps and 2853 refills, we retrospectively analyzed the actual flow rate from 09/2010 to 09/2018. A relevant flow rate variance is evident after three years; therefore, flow rate monitoring and dose adjustment are mandatory.
Citation Styles
Harvard Citation style: Richter, M., Harutyunova, S., Bollmann, T., Classen, S., Fuge, J., Gall, H., et al. (2020) Flow rate variance of a fully implantable pump for the delivery of intravenous treprostinil in pulmonary arterial hypertension, Pulmonary Circulation, 10(1), Article 2045894020910136. https://doi.org/10.1177/2045894020910136
APA Citation style: Richter, M., Harutyunova, S., Bollmann, T., Classen, S., Fuge, J., Gall, H., Gerhardt, F., Ghofrani, H., Gunkel, H., Gruenig, E., Halank, M., Heine, A., Klose, H., Lange, T., Neurohr, C., Nickolaus, K., Opitz, C., Rosenkranz, S., Seyfarth, H., ...Olsson, K. (2020). Flow rate variance of a fully implantable pump for the delivery of intravenous treprostinil in pulmonary arterial hypertension. Pulmonary Circulation. 10(1), Article 2045894020910136. https://doi.org/10.1177/2045894020910136
Keywords
implantable infusion pump; prostanoid; Pulmonary hypertension