Journal article
Authors list: Schmidt, Alexander; Billig, Jan-Wilhelm; Schlenz, Maximiliane A.; Rehmann, Peter; Wostmann, Bernd
Publication year: 2019
Pages: 430-432
Journal: International Journal of Prosthodontics
Volume number: 32
Issue number: 5
ISSN: 0893-2174
eISSN: 1942-4426
DOI Link: https://doi.org/10.11607/ijp.6371
Publisher: Quintessence Publishing
Abstract:
Purpose: To assess the manufacturing tolerances of three different intraoral scanbodies (ISBs) and their possible influence on the transfer of the intraoral implant position to the digital cast. Materials and Methods: Three different ISBs were measured using x-ray computed tomography. The length, diameter, angle, and target volume deviation were analyzed, and the data were subjected to a univariate analysis of variance. Results: There were significant differences between all three ISBs in length, diameter, and target volume deviations (P < .05). Conclusion: Manufacturing tolerances of ISBs have the potential to decisively affect the transfer accuracy of the implant position from the recorded intraoral position to the digital cast, which needs to be considered.
Citation Styles
Harvard Citation style: Schmidt, A., Billig, J., Schlenz, M., Rehmann, P. and Wostmann, B. (2019) Influence of the Accuracy of Intraoral Scanbodies on Implant Position: Differences in Manufacturing Tolerances, International Journal of Prosthodontics, 32(5), Article PMID 31486814. pp. 430-432. https://doi.org/10.11607/ijp.6371
APA Citation style: Schmidt, A., Billig, J., Schlenz, M., Rehmann, P., & Wostmann, B. (2019). Influence of the Accuracy of Intraoral Scanbodies on Implant Position: Differences in Manufacturing Tolerances. International Journal of Prosthodontics. 32(5), Article PMID 31486814, 430-432. https://doi.org/10.11607/ijp.6371