Journalartikel
Autorenliste: Scholz, P. A.; Kraft-Bermuth, S.; Andrianov, V.
Jahr der Veröffentlichung: 2016
Seiten: 576-582
Zeitschrift: Journal of Low Temperature Physics
Bandnummer: 184
Heftnummer: 3-4
ISSN: 0022-2291
eISSN: 1573-7357
DOI Link: https://doi.org/10.1007/s10909-015-1438-z
Verlag: Springer
Abstract:
High-precision X-ray spectroscopy of highly charged ions at storage rings provides a sensitive test of quantum electrodynamics in strong Coulomb fields. To increase the precision of such experiments, silicon microcalorimeters have already been applied successfully. To minimize the interruption of beam times due to maintenance, a new cryogen-free He/He dilution refrigerator has been designed and is under commissioning. However, in cryogen-free systems microphonic noise due to vibrations contributes considerably to the overall noise and may limit the detector energy resolution. Therefore, we report on systematic vibration studies on a cryogen-free He/He dilution refrigerator which is specially adapted for experiments at storage rings.
Zitierstile
Harvard-Zitierstil: Scholz, P., Kraft-Bermuth, S. and Andrianov, V. (2016) Systematic Vibration Studies on a Cryogen-Free He/He Dilution Refrigerator for X-ray Spectroscopy at Storage Rings, Journal of Low Temperature Physics, 184(3-4), pp. 576-582. https://doi.org/10.1007/s10909-015-1438-z
APA-Zitierstil: Scholz, P., Kraft-Bermuth, S., & Andrianov, V. (2016). Systematic Vibration Studies on a Cryogen-Free He/He Dilution Refrigerator for X-ray Spectroscopy at Storage Rings. Journal of Low Temperature Physics. 184(3-4), 576-582. https://doi.org/10.1007/s10909-015-1438-z
Schlagwörter
HEAVY-IONS; LOW-TEMPERATURE DETECTORS; Microcalorimeters for X-rays; Reduction of microphonics; Vibration analysis of pulse tube coolers