Journalartikel

Intra- and Interobsever Reliability of Sonographic Assessment of the Optic Nerve Sheath Diameter in Healthy Adults


AutorenlisteBaeuerle, Jochen; Lochner, Piergiorgio; Kaps, Manfred; Nedelmann, Max

Jahr der Veröffentlichung2012

Seiten42-45

ZeitschriftJournal of Neuroimaging

Bandnummer22

Heftnummer1

ISSN1051-2284

eISSN1552-6569

DOI Linkhttps://doi.org/10.1111/j.1552-6569.2010.00546.x

VerlagWiley


Abstract

BACKGROUND

The use of an ultrasound-based evaluation of the optic nerve sheath diameter (ONSD) has previously been demonstrated for detecting raised intracranial pressure. In order to be feasible in clinical workup, the test qualities of transorbital ultrasonography need to be determined. The aim of this study was therefore to establish normal values and to assess the intra-and interobserver reliability of this method.

METHODS

Using a 9-3 MHz linear array transducer, the ONSD of 40 healthy subjects was independently measured by 2 investigators.

RESULTS

Depicting the optic nerve and its sheath was possible in all individuals. The mean ONSD was 5.4 +/- .6 mm with a range of 4.3-7.6 mm. The intraobserver reliability analyzed with Cronbach's Alpha was found to be high with values between .92 and .97. Pearson's correlation coefficient between the 2 investigators was .81 on the right side and .84 on the left. There was no correlation between ONSD and age, body mass index, or gender.

CONCLUSIONS

Transorbital B-mode sonography is a feasible method to assess the ONSD with a high intra- and interobsever reliability. Normal values on ONSD are presented in this study that will be useful in future studies on pathological conditions.




Zitierstile

Harvard-ZitierstilBaeuerle, J., Lochner, P., Kaps, M. and Nedelmann, M. (2012) Intra- and Interobsever Reliability of Sonographic Assessment of the Optic Nerve Sheath Diameter in Healthy Adults, Journal of Neuroimaging, 22(1), pp. 42-45. https://doi.org/10.1111/j.1552-6569.2010.00546.x

APA-ZitierstilBaeuerle, J., Lochner, P., Kaps, M., & Nedelmann, M. (2012). Intra- and Interobsever Reliability of Sonographic Assessment of the Optic Nerve Sheath Diameter in Healthy Adults. Journal of Neuroimaging. 22(1), 42-45. https://doi.org/10.1111/j.1552-6569.2010.00546.x



Schlagwörter


FUNDAMENTALSintra- and interobserver reliabilityINTRACRANIAL HYPERTENSIONOptic nerve sheath diameterTransorbital sonography


Nachhaltigkeitsbezüge


Zuletzt aktualisiert 2025-02-04 um 02:47