Journal article

Influence of impression technique and material on the accuracy of multiple implant impressions


Authors listWoestmann, Bernd; Rehmann, Peter; Balkenhol, Markus

Publication year2008

Pages299-301

JournalInternational Journal of Prosthodontics

Volume number21

Issue number4

ISSN0893-2174

PublisherQuintessence Publishing


Abstract
This study aimed to analyze the influence of impression technique (pick-up versus reposition) and material on the accuracy of the working cast. Sixty impressions were made with 3 materials from a master cast with 4 XiVE implants. The changes in the implant axis direction, rotation, and 3-dimensional shift were assessed. The pick-up technique showed significantly (P .05, U test) lower values for axis direction and 3D shift but higher values for rotation than the reposition technique. The differences between the materials were not significant (P > .05, H test). It can be concluded that the impression technique-in contrast to the impression material-has a significant influence on transfer accuracy.



Citation Styles

Harvard Citation styleWoestmann, B., Rehmann, P. and Balkenhol, M. (2008) Influence of impression technique and material on the accuracy of multiple implant impressions, International Journal of Prosthodontics, 21(4), pp. 299-301

APA Citation styleWoestmann, B., Rehmann, P., & Balkenhol, M. (2008). Influence of impression technique and material on the accuracy of multiple implant impressions. International Journal of Prosthodontics. 21(4), 299-301.