Precision energy measurement of the 166 keV γ ray in 139Ce decay


Authors listAlburger, DE; Wesselborg, C

Publication year1999

Pages49-51

JournalNuclear Instruments and Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors and Associated Equipment

Volume number423

Issue number1

ISSN0168-9002

DOI Linkhttps://doi.org/10.1016/S0168-9002(98)01177-2

PublisherElsevier


Abstract
An intrinsic Ge (LEPS-type) detector was used to measure the energy of the 166 keV gamma ray in Ce-139 decay against the gamma rays of Ta-182. Reference values for Ta-182 were taken from the recent list of gamma ray energy standards as updated by Helmer and van der Leun. Based on 18 runs taken under a variety of conditions, the average value for this La-139 gamma ray is 165.8575(11) keV. This is mid-way between two earlier results, but about six times as accurate. Some advantages of 139Ce as both a gamma ray and internal-conversion electron calibration source are pointed out. (C) 1999 Elsevier Science B.V. All rights reserved.



Citation Styles

Harvard Citation styleAlburger, D. and Wesselborg, C. (1999) Precision energy measurement of the 166 keV γ ray in 139Ce decay, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 423(1), pp. 49-51. https://doi.org/10.1016/S0168-9002(98)01177-2

APA Citation styleAlburger, D., & Wesselborg, C. (1999). Precision energy measurement of the 166 keV γ ray in 139Ce decay. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 423(1), 49-51. https://doi.org/10.1016/S0168-9002(98)01177-2


Last updated on 2025-02-04 at 06:06