PRECISION ENERGY MEASUREMENTS OF GAMMA-RAYS FROM TI-44 AND BA-133


AutorenlisteWESSELBORG, C; ALBURGER, DE

Jahr der Veröffentlichung1991

Seiten89-90

ZeitschriftNuclear Instruments and Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors and Associated Equipment

Bandnummer302

Heftnummer1

ISSN0168-9002

DOI Linkhttps://doi.org/10.1016/0168-9002(91)90496-D

VerlagElsevier


Abstract
A LEPS-type Ge detector was used for precision energy measurements of the 68 and 78 keV gamma rays from the decay of Ti-44, with Ba-133 and Ta-182 as reference sources. The values obtained were 67.8679(14) keV and 78.3234(10) keV. The latter value is 13.6 eV lower than a previous report and suggests that systematic effects were present in the earlier determination of the 78 keV energy. The Ba-133 81 keV line was measured to be 80.9964(21) keV and we adopt 80.991(14) keV from the weighted average of all reported results.



Zitierstile

Harvard-ZitierstilWESSELBORG, C. and ALBURGER, D. (1991) PRECISION ENERGY MEASUREMENTS OF GAMMA-RAYS FROM TI-44 AND BA-133, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 302(1), pp. 89-90. https://doi.org/10.1016/0168-9002(91)90496-D

APA-ZitierstilWESSELBORG, C., & ALBURGER, D. (1991). PRECISION ENERGY MEASUREMENTS OF GAMMA-RAYS FROM TI-44 AND BA-133. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 302(1), 89-90. https://doi.org/10.1016/0168-9002(91)90496-D