Zeitschrift: Nuclear Instruments and Methods in Physics Research. Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Bandnummer: 423
Heftnummer: 1
ISSN: 0168-9002
DOI Link: https://doi.org/10.1016/S0168-9002(98)01177-2
Verlag: Elsevier
Abstract:
An intrinsic Ge (LEPS-type) detector was used to measure the energy of the 166 keV gamma ray in Ce-139 decay against the gamma rays of Ta-182. Reference values for Ta-182 were taken from the recent list of gamma ray energy standards as updated by Helmer and van der Leun. Based on 18 runs taken under a variety of conditions, the average value for this La-139 gamma ray is 165.8575(11) keV. This is mid-way between two earlier results, but about six times as accurate. Some advantages of 139Ce as both a gamma ray and internal-conversion electron calibration source are pointed out. (C) 1999 Elsevier Science B.V. All rights reserved.
Zitierstile
Harvard-Zitierstil: Alburger, D. and Wesselborg, C. (1999) Precision energy measurement of the 166 keV γ ray in 139Ce decay, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 423(1), pp. 49-51. https://doi.org/10.1016/S0168-9002(98)01177-2
APA-Zitierstil: Alburger, D., & Wesselborg, C. (1999). Precision energy measurement of the 166 keV γ ray in 139Ce decay. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 423(1), 49-51. https://doi.org/10.1016/S0168-9002(98)01177-2